SmartLog V4 Troubleshooting Guide
Why doesn’t the SmartLog V4 power up?
Will the SmartLog V4 log test results without the TEAM software?
How do I set the SmartLog V4 to test ESD shoes?
Why do I receive a fail lo when testing my ESD footwear?
Why does the SmartLog V4 chirp several times when passing tests?
Where can I download the TEAM software?
Where is the License Key required to access the TEAM software?
Why does the software require a license key when the USB dongle is plugged in?
How can I connect the SmartLog V4 if I have no available serial ports?
The host computer is not recognizing the SmartLog V4?
Why does the SmartLog V4 lose signal with the TEAM Software?
Are the data loggers compatible with the TEAM software?
Will the older SmartLog V4’s (RS-485 jack) daisy chain with the 2-pin connector?
What is the maximum distance the SmartLog X3’s can be daisy chained?
Why do the SmartLog V4’s that are daisy-chained lose signal?
Why do I receive an error when setting up my E-mail?
How can I customize my own display message in the TEAM software?
Why doesn’t the SmartLog V4 power up?
- Verify the incoming voltage meets the power requirements
- Verify the power is connected to the back of the clock labeled “12VDC IN” and not the
tester
Go back to top
Will the SmartLog V4 log test results without the TEAM software?
- Yes, the BC Reader (clock) will store up to 2,500 tests and hold up to 10,000 user id’s. The tests can then be pulled into the TEAM software when Polling is enabled.
Go back to top
How do I set the SmartLog V4 to test ESD shoes?
- ESD shoes have a tendency of testing in the 10E8 – 10E9 range. The Fail Hi on the
SmartLog X3 is set to 35Meg (10E7).
We recommend adjusting the footwear Hi settings
to 1Gig. (Refer to page 4 of TB-6564)
Go back to top
Why do I receive a fail lo when testing my ESD footwear?
- Verify the ESD footwear being tested is within spec
- Verify enough moisture is being generated from the user
- Verify the stereo plug is securely fastened into the SmartLog V4 tester and footplate
Go back to top
Why does the SmartLog V4 chirp several times when passing tests?
- The chirping is the door relay. To turn this feature off or reduce the amount of chirps, go
to the “Test Machine” tab of the TEAM software and modify the “Door Relay.”
Go back to top
Where can I download the TEAM software?
- The TEAM software installation CD is included with the initial purchase of the SmartLog V4
- The entire TEAM software and upgrades can be downloaded from the DescoEMIT.com website.
Go back to top
Where is the License Key that is required to access the TEAM software?
- A Green USB Dongle is included with the purchase of the TEAM5 or TEAM5 Enterprise software. It must be plugged in to an available active
USB Port of the host computer where the software is installed and accessed.
- If you cannot locate the USB Dongle, the factory will provide you with a temporary
license until a new key is purchased.
Go back to top
Why does the software require a license key when the USB dongle is plugged in?
- Insert the Green USB dongle into another available USB port.
- The driver used to access the dongle may have been removed or misplaced. To re-install
the drive, refer to the attached link (http://cscsoftware.net/usb.exe).
Go back to top
How can I connect the SmartLog V4 if I have no available serial ports?
- The SmartLog V4 can be connected via an Ethernet Adapter
- The SmartLog V4 can be connected to a USB port using a serial to USB converter
Go back to top
The host computer is not recognizing the SmartLog V4?
- Verify a flat non-inverted cable is being used. When the correct cable is used, the two
LEDs at the bottom of the SmartLog V4 clock will blink during transmission
- Verify the cable is connected to the RS-232 port and not the RS-485 port
- Verify the SmartLog V4 is set to Master and not Slave on the back of the BC Reader
- Verify the Baud Rate of both the SmartLog V4 and TEAM software is set to 9600
- Verify the com port that’s designated to the SmartLog V4 is active and working properly
- Verify the TEAM software is set to locate the designated com port. Go to the “Test Machine” tab and increase the Max Com Port to 10. The software will scan all active com ports to locate the SmartLog V4. If the software is still no recognizing the SmartLog V4,
try another com port.
Go back to top
Why does the SmartLog V4 lose signal with the TEAM Software?
- Verify the Baud Rate of both the SmartLog V4 and TEAM software is set to 9600
- The Wait (sec) or network wait time can be increased up to 5 seconds to compensate for network lag time. A lower wait time of less than 1 sec. is recommended for SmartLogs that are directly connected to a com port.
- Some windows processes like firewalls and virus scanners may prevent TEAM or the Digi Virtual com port software from connecting to Smartlog V4’s. Contact your IT department
regarding these programs.
- Each Smartlog V4 must have a unique network address. Although the address is stored in
Non-Volatile memory it is possible for the address to reset after a power down/up cycle.
Go back to top
Are the data loggers compatible with the TEAM software?
- No, the data logger uses older firmware that is not compatible with TEAM.
Go back to top
Will the older SmartLog V4’s (RS-485 jack) daisy chain with the 2-pin connector?
- Yes, contact the factory for special instructions.
Go back to top
What is the maximum distance the SmartLog V4’s can be daisy chained?
- Maximum distance from the host computer to the SmartLog V4 is 50ft
- Maximum distance from the last SmartLog V4 to the first is 200ft
Go back to top
Why do the SmartLog V4’s that are daisy-chained lose signal?
- Verify each additional SmartLog V4 has a different set of ID’s. (Refer to TB-6547)
- Verify the SmartLog V4’s are not near any heavy operating machinery
- Verify the correct cable is being used and connected to the RS-485 port
- Verify the distance between each SmartLog V4 does not exceed 50 ft
- Verify the distance from the first to the last SmartLog V4 does not exceed 200 ft
Go back to top
Why do I receive an error when setting up my E-mail?
-
Verify your Firewall or Virus Scan programs are not preventing the e-mail setup
-
Verify your e-mail server accepts SMTP mail
-
Verify your e-mail server does not require an ID or password
-
Verify your e-mail connection(s) match the TEAM software
Go back to top
How can I customize my own display message in the TEAM software?
- Turn Polling off. Under the Test Machine tab in the Display Language window, type in the
desired message for each Title. Select the save icon and turn polling back on.
Go back to top